The 500 Hz tone-pip evoked ABRs are below.
- Where would you mark wave V for the left ear traces? Keep in mind that wave V latency will prolong as you decrease intensity. What is/are the role(s) for latency in tone-pip threshold estimation?
The audiologist did not run a control run, but subtracted the initial and replication runs to give an approximation of the control run. That is shown on the right bottom.
- Which right ear response(s) would you mark as being consistent with a present response? Compare latencies between ears.